Magnetic Force Microscopy for Nanoparticle Characterization

Gustavo Cordova1, Brenda Yasie Lee1 and Zoya Leonenko1,2,3*

1Department of Biology, University of Waterloo, Waterloo ON, Canada
2Department of Physics and Astronomy, University of Waterloo, Waterloo ON, Canada
3Waterloo Institute for Nanotechnology, University of Waterloo, Waterloo ON, Canada

Abstract

Since the invention of the atomic force microscope (AFM) in 1986, there has been a drive to apply this scanning probe technique or a form of this technique to various disciplines in nanoscale science. Magnetic force microscopy (MFM) is a member of a growing family of scanning probe methods and has been widely used for the study of magnetic materials. In MFM a magnetic probe is used to raster-scan the surface of the sample, of which its magnetic field interacts with the magnetic tip to offer insight into its magnetic properties. This review will focus on the use of MFM in relation to nanoparticle characterization, including superparamagnetic iron oxide nanoparticles, covering MFM imaging in air and in liquid environments.

Keywords

Atomic force microscopy, Magnetic force microscopy (MFM), Nanoparticles, MFM Imaging in air and in liquid, Magnetic and superparamagnetic iron oxide nanoparticles (SPIONs)

Abbreviations

AFM: Atomic Force Microscopy; HL-60: Human Promyelocytic Leukemia Cells; MCF-7: Michigan Cancer Foundation-7 (breast cancer cell line); MFM: Magnetic Force Microscopy; PMMA: Poly(methyl methacrylate); PS: Pulmonary Surfactant; SMNP: Supramolecular Magnetic Nanoparticles; SPION: Superparamagnetic Iron Oxide Nanoparticles; SQUID: Superconducting Quantum Interference Device

Introduction

Magnetic force microscopy (MFM) is a type of scanning probe microscopy where a magnetic probe is used to scan the sample surface while interacting with the magnetic fields of the sample [1, 2]. The most common method is called the “Two- Pass Technique” or “Hover Mode Scanning Method” [3-5], in which the sample is scanned twice (see Figure 1): once to produce a topographical image (AFM) and a second time to produce an MFM image. This is the most commonly used method for MFM, because it is relatively simple to implement and will provide high phase contrast in ambient condition [6].

On the first trace of the sample, the cantilever directly scans the surface of the sample just as it would for AFM in its intermittent contact or “tapping” mode, producing a topographical image. During this time, short range interactions (e.g. Van der Waals forces) have the most significant effect on the cantilever. Subsequently on the retrace, the cantilever is raised to a user-defined height (eg. 50 nm) away from the surface and scans for the magnetic signal by following the topographical pattern from the previous trace as seen in Figure 1A. Long range interactions, such as magnetic forces, are the most prevalent in the retrace and thus allow the MFM image to reflect the magnetic properties of the sample being imaged.

Theory and Advances

Magnetic force microscopy was developed in 1987 as a new method for imaging magnetic fields down to a resolution of 100 nm [7]. This novel method used the magnetic force gradient acting between a magnetic tip and magnetic sample surface to map the magnetic field distributions on the nanoscale. The instrumentation setup involved a 25 μm diameter iron wire which served as both the cantilever and the tip, where one side of the wire was tapered down to a diameter of 2 μm with a 0.1 nm diameter tip etched at its end. The resolution achieved was 100 nm. Although the operating principle of MFM is identical to AFM, the interactions between the tip and the sample are of magnetic nature and are different from forces acting in AFM.

MFM measurements are commonly taken in the Two- Pass Technique, where AFM topography is measured in the first pass, and the second pass involves the retrace of the first pass at a small z-offset to record long-range magnetic forces. As the tip oscillates, a decaying magnetic force is experienced as the tip-sample separation increases; it is this force gradient that changes the phase of the cantilever’s oscillation and allows for a phase signal to be recorded during the MFM measurement. The phase signal (ξ) can be calculated using the following formula, which is derived from the application of the Euler beam theory to the force gradient (fm) and phase shift (Q·f ’m·k-1) [8]:

Where γ1 = 0.97, a constant that originates from the analysis of the thermal motion of a cantilever [9].

Due to its obvious application to magnetic materials research, MFM became a popular method in this field. However, there were several shortcomings during its early years, specifically with regards to its resolution and sensitivity to achieve measurements below 100 nm [10]. In order to improve the spatial resolution and signal strength of the MFM technique, different methods and advancements were tested to optimize the magnetized portion of the probe, the tipsample distance, and the sample surface. Techniques emerged to fabricate advanced MFM tips or probes with tip sizes as small as 5 nm in diameter using electron beam deposition, focused ion beam milling, and attaching fullerene carbon multi walled nanotubes. Minimizing the size of the force-sensitive component of the probe (the tip apex) and keeping the tip close to the sample surface would help to improve lateral resolution [10]. Even with increased resolution, the interpretation of MFM signals and images is complex and requires further understanding of the types of interactions occurring between the tip and the sample. A recent study was performed to determine which sources of interaction contributed the most to the MFM signal, and it was discovered that the observed MFM response was mainly from electrostatic interactions and not with the magnetism of the sample itself [11]. As a result, correlating the MFM signal with the magnetism of the sample will require further extensive analysis of the different interactions that play a role between the tip and sample.

The bimodal AFM technique has been used to corroborate the theory proposed by Rodriguez and Garcia that adding a second resonant mode with a bimodally excited cantilever would be highly sensitive to long-range forces such as van der Waals interactions [12, 13]. This bimodal MFM method allows for one-pass topographical and magnetic force imaging with high spatial resolution, with image quality equivalent to that of the standard two-pass MFM. This technique has also been combined with the use of an external magnetic field to excite the magnetic tip of the probe to detect various magnetic materials. It was demonstrated that its sensitivity and signalto- noise ratio were superior to conventional MFM techniques [14]. In both studies, it was observed that this technique also had increased spatial resolution when the tip was closer to the sample surface [12, 14].

Advantages and Applications

MFM offers certain advantages over other imaging techniques, which has led to its increasing prevalence as a tool for magnetic nanoparticle characterization [15-20]. It has the capability to separate the magnetic interactions from the other tip sample forces (such as van der Waals, and other forces recorded in AFM). MFM can be used to detect nanoscale magnetic domains as well as simultaneously obtain both AFM phase and topography images, something that is not possible using more traditional nanoscale imaging techniques such as transmission or scanning electron microscopy. In our recent study [3], we examined the application of MFM for the detection and localization of superparamagnetic iron oxide nanoparticles (SPIONs). Figure 2 shows the corresponding MFM phase and topographical images for bare as well as silica-coated SPIONs. A study in the Universitat de Barcelona in Spain demonstrated that MFM could be used to distinguish between superparamagnetic and blocked states in aggregates of iron oxide nanoparticles [21]. Another study used MFM to investigate the magnetic microstructure of nanoparticles with single- and multi-domain octahedral particles [22].

Perhaps most notably, MFM is capable of operating under ambient conditions, at varying temperatures, ultrahigh vacuums, as well as liquid environments [3, 23] while still providing resolution to less than ten nanometers. This is essential because it allows magnetic nanoparticles to be localized and characterized in vitro [24] or inside polymer films and cell-like systems [3, 25, 26]. Because of the pervasiveness of magnetic nanoparticles in biomedical applications, it is crucial that when these nanoparticles are characterized, it is done so under conditions relevant to their application (e.g. a physiologically relevant environment, most importantly inside the cells). Until recently, the potential of MFM in this regard has been largely unexplored. Figures 3 and 4 shows separate studies where various cell lines were incubated with magnetic nanoparticles, both imaged using MFM in ambient conditions.

Liquid imaging of SPION nanoparticles was also reported by Cordova et al. in 2014 [3] using the Two-Pass Technique (Figure 5). Until this point, the majority of MFM imaging of magnetic nanoparticles had only been undertaken in ambient conditions. The single-pass bimodal MFM technique had also been successfully used to detect and image superparamagnetic nanoparticles in liquid down to 5 nm [27]. Imaging magnetic nanoparticles in liquid served as a proof of principle for using MFM to image magnetic nanoparticles within cells in order to better understand their distribution and behavior in a physiologically relevant environment. Another recent study published from the Universidad Autonoma de Madrid in Spain reported successful MFM imaging of nanoparticles in liquid [23]. In this study, MFM was used to not only image these nanoparticles, but to optimize the MFM signal acquisition in liquid media so that it could be applied to the characterization of magnetic nanoparticles and their magnetic signals. They were able to image the magnetic domains of hard disk drives as well as DMSA-coated Fe3O4 ferromagnetic nanoparticles in both air and liquid environments (Figure 6) [23]. To maximize the MFM signal in drive amplitude modulation mode for AFM, the group varied relevant measuring and operating conditions such as cantilever oscillation amplitude, set point, and Z lift distance. The authors [23], observed that the magnetic signal decreased with tip-sample distance, as opposed to previous study [3], where slight increase was observed within the distances studied. This discrepancy may originate due to coupling between the electrical and magnetic signal. More research will need to be performed in relation to MFM imaging in liquid and on effect of tip-sample distance on the cross-coupling between the magnetic and electrical signals.

In terms of biological applications, MFM has also been used in studying the magnetic properties of biomolecules, magnetotactic bacterium [28] and biogenic nanoparticles [29]. Magnetic nanoparticles naturally occurring in biological systems have also been studied, including iron compounds associated with neurological disorders [30], magnetic domains in magnetotactic bacteria [31] and iron deposits in livers with Hepatitis B [32].

Moreover, MFM allows the user to characterize magnetic nanoparticles in a non destructive manner without any special surface preparation or nanoparticle modification. In addition, MFM allows one to determine the magnetic moment of a single nanoparticle and how this measurement changes with nanoparticle size as well as probe distance from the sample, something which bulk magnetic analysis is not capable of. As all scanning probe methods it provides high resolution imaging of nanoparticle topography and quantitative analysis of nanoparticle size and density distribution, when nanoparticles are deposited on solid substrates. With successful development of MFM operating in liquid it opens excellent possibilities of studying magnetic nanoparticles in biologically relevant conditions, such as inside the cells, where AFM when AFM imaging is not working. Considering rapid development of novel applications of magnetic nanoparticles in medicine and biomedical nanotechnology, as therapeutic agents, contrast agents in MRI imaging and drug delivery [3] MFM characterization of nanoparticles becomes more valuable and desirable. Overall, MFM has proven itself to be an effective yet underused tool that offers great potential for the localization and characterization of magnetic nanoparticles.

Conflict of Interest

The authors declare that they have no conflicts of interest with the contents of this article.

Authors Contribution

Gustavo Cordova and Brenda Yasie Lee are contributed equally to this work.

Acknowledgments

We greatly acknowledge the funding from Natural Science and Engineering Research Council of Canada (NSERC).

References

1. de Lozanne A. 2006. Application of magnetic force microscopy in nanomaterials characterization. Microsc Res Tech 69(7): 550-562. doi: 10.1002/jemt.20325

2. Karoutsos V. 2009. Scanning probe microscopy: instrumentation and applications on thin films and magnetic multilayers. J Nanosci Nanotechnol 9(12): 6783-6798.  doi: 10.1166/jnn.2009.1474

3. Cordova G, Attwood S, Gaikwad R, Gu F, Leonenko Z. 2014. Magnetic force microscopy characterization of superparamagnetic iron oxide nanoparticles (SPIONs). Nano Biomed Eng 6(1): 1-12. doi: 10.5101/nbe.v6i1.p31-39

4. Hsieh CW, Zheng B, Hsieh S. 2010. Ferritin protein imaging and detection by magnetic force microscopy. Chem Commun (Camb) 46(10): 1655-1657. doi: 10.1039/B912338E

5. Torre B, Bertoni G, Fragouli D, Falqui A, Salerno M, et al. 2011. Magnetic force microscopy and energy loss imaging of superparamagnetic iron oxide nanoparticles. Sci Rep 1: 202. doi: 10.1038/srep00202

6. Neves CS, Quaresma P, Baptista PV, Carvalho PA, Araújo JP. 2010. New insights into the use of magnetic force microscopy to discriminate between magnetic and nonmagnetic nanoparticles. Nanotechnology 21(30): 305706. doi: 10.1088/0957-4484/21/30/305706

7. Martin Y, Wickramasingh, HK. 1987. Magnetic imaging by ‘‘force microscopy’’ with 1000 Å resolution. Applied Physics Letters 50(20): 1455-1457. doi: 10.1063/1.97800

8. Colin R, Colm D. 2012. Performing quantitative MFM measurements on soft magnetic nanostructures. Nanotechnology 23(45): 455701. doi: 10.1088/0957-4484/23/45/455701

9. Cook SM, Schäffer TE, Chynoweth KM, Wigton M, Simmonds RW, et al. 2006. Practical implementation of dynamic methods for measuring atomic force microscope cantilever spring constants. Nanotechnology 17(9): 2135. doi: 10.1088/0957-4484/17/9/010

10. Koblischka MR, Hartmann U. 2003. Recent advances in magnetic force microscopy. Ultramicroscopy 97(1-4): 103-112. doi: 10.1016/S0304- 3991(03)00034-2

11. Li LH, Chen Y. 2014. Electric contributions to magnetic force microscopy response from graphene and MoS2 nanosheets. Journal of Applied Physics 116: 213904. doi: 10.1063/1.4903040

12. Li JW, Cleveland JP, Proksch R. 2009. Bimodal magnetic force microscopy: Separation of short and long range forces. Applied Physics Letters 94(16): 163118. doi: 10.1063/1.3126521

13. Rodrı́guez TR, Garcı́a R. 2004. Compositional mapping of surfaces in atomic force microscopy by excitation of the second normal mode of the microcantilever. Applied Physics Letters 84(3): 449-451. doi: 10.1063/1.1642273

14. Liu D, Mo K, Ding X, Zhao L, Lin G, et al. 2015. Secondary resonance magnetic force microscopy using an external magnetic field for characterization of magnetic thin films. Applied Physics Letters 107(10): 103110. doi: 10.1063/1.4930878

15. Park JW, Yoo IS, Chang WS, Lee EC, Ju H, et al. 2008. Magnetic moment measurement of magnetic nanoparticles using atomic force microscopy. Measurement Science and Technology 19(1): 017005. doi: 10.1088/0957-0233/19/1/017005

16. Pedreschi F, Sturm JM, O’Mahony JD, Flipse CFJ. 2003. Magnetic force microscopy and simulations of colloidal iron nanoparticles. Journal of Applied Physics 94(5): 3446-3450. doi: 10.1063/1.1593219

17. Puntes VF, Gorostiza P, Aruguete DM, Bastus NG, Alivisatos AP. 2004. Collective behaviour in two-dimensional cobalt nanoparticle assemblies observed by magnetic force microscopy. Nature Materials 3(4): 263-268. doi: 10.1038/nmat1094

18. Ramlan DG, May SJ, Zheng JG, Allen JE, Wessels BW, et al. 2005. Ferromagnetic self-assembled quantum dots on semiconductor nanowires. Nano Letters 6(1): 50-54. doi: 10.1021/nl0519276

19. Sievers S, Albrecht M, Siegner U, Herweg C, Freyhardt HC. 2005. Self-ordered growth and magnetic force microscopy study of iron nanoparticles. Journal of Applied Physics 97: 10J308. doi: 10.1063/1.1851927

20. Takamura Y, Chopdekar RV, Scholl A, Doran A, Liddle JA, et al. 2006. Tuning magnetic domain structure in nanoscale La0.7Sr0.3MnO3 islands. Nano Letters 6(6): 1287-1291. doi: 10.1021/nl060615f

21. Moya C, Iglesias-Freire Ó, Batlle X, Labarta A, Asenjo A. 2015. Superparamagnetic versus blocked states in aggregates of Fe3-xO4 nanoparticles studied by MFM. Nanoscale 7(42): 17764-17770. doi: 10.1039/C5NR04424C

22. Datt G, Sen Bishwas M, Manivel Raja M, Abhyankar AC. 2016. Observation of magnetic anomalies in one-step solvothermally synthesized nickel-cobalt ferrite nanoparticles. Nanoscale 8(9): 5200-5213. doi: 10.1039/C5NR06791J

23. Ares P, Jaafar M, Gil A, Gómez-Herrero J, Asenjo A. 2015. Magnetic force microscopy in liquids. Small 11(36): 4731-4736. doi: 10.1002/smll.201500874

24. Raşa M, Kuipers BW, Philipse AP. 2002. Atomic force microscopy and magnetic force microscopy study of model colloids. Journal of Colloid and Interface Science 250(2): 303-315. doi: 10.1006/jcis.2002.8345

25. Shen HB, Long DH, Zhu LZ, Li XY, Dong YM, et al. 2006. Magnetic force microscopy analysis of apoptosis of HL-60 cells induced by complex of antisense oligonucleotides and magnetic nanoparticles. Biophys Chem 122(1): 1-4. doi: 10.1016/j.bpc.2006.01.003

26. Zhang Y, Yang M, Ozkan M, Ozkan CS. 2009. Magnetic force microscopy of iron oxide nanoparticles and their cellular uptake. Biotechnol Prog 25(4): 923-928. doi: 10.1002/btpr.215

27. Dietz C, Herruzo ET, Lozano JR, Garcia R. 2011. Nanomechanical coupling enables detection and imaging of 5 nm superparamagnetic particles in liquid. Nanotechnology 22(12): 125708. doi: 10.1088/0957-4484/22/12/125708

28. Proksch RB, Schäffer TE, Moskowitz BM, Dahlberg ED, Bazylinski DA, et al. 1995. Magnetic force microscopy of the submicron magnetic assembly in a magnetotactic bacterium. Applied Physics Letters 66(19): 2582-2584. doi: 10.1063/1.113508

29. Albrecht M, Janke V, Sievers S, Siegner U, Schüler D, et al. 2005. Scanning force microspy study of biogenic nanoparticles for medical applications. Journal of Magnetism and Magnetic Materials 290-291(Part 1): 269-271. doi: 10.1016/j.jmmm.2004.11.206

30. Dobson J. 2001. Nanoscale biogenic iron oxides and neurodegenerative disease. FEBS Letters 496(1): 1-5. doi: 10.1016/S0014-5793(01)02386-9

31. Diebel CE, Proksch R, Green CR, Neilson P, Walker MM. 2000. Magnetite defines a vertebrate magnetoreceptor. Nature 406(6793): 299-302. doi: 10.1038/35018561

32. Martinelli AL, Filho AB, Franco RF, Tavella MH, Ramalho LN, et al. 2004. Liver iron deposits in hepatitis B patients: association with severity of liver disease but not with hemochromatosis gene mutations. J Gastroenterol Hepatol 19(9): 1036-1041. doi: 10.1111/j.1440-1746.2004.03410.x

*Correspondence to:

Dr. Zoya Leonenko
Department of Physics and Astronomy,
University of Waterloo,
200 University Avenue West, Waterloo,
Ontario, Canada
Tel: 519-888-4567 x 38273
E-mail: zleonenk@uwaterloo.ca

Received: March 04, 2016
Accepted: April 25, 2016
Published: April 28, 2016

Citation: Cordova G, Lee BY, Leonenko Z. 2016. Magnetic Force Microscopy for Nanoparticle Characterization. NanoWorld J 2(1): 10-14.

Copyright: © 2016 Cordova et al. This is an Open Access article distributed under the terms of the Creative Commons Attribution 4.0 International License (CC-BY) (http://creativecommons.org/licenses/by/4.0/) which permits commercial use, including reproduction, adaptation, and distribution of the article provided the original author and source are credited.

Published by United Scientific Group
© NanoWorld Journal 2016 | United Scientific Group, #8105, Suite 112, Rasor Blvd, PLANO, TX 75024, USA | Terms and Conditions | Privacy Policy